Diffusion-Weighted Imaging of Sellar Region: A Comparison Study of BLADE sequence and Single-Shot Echo Planar Imaging sequence

Yiping Lu 1 , Bo Yin 2 , and Daoying Geng 2

1 Radiology Department, Huashan Hospital affiliated to Fudan University, Shanghai, Shanghai, China, 2 Radiology Department, Huashan Hospital affiliated to Fudan University, Shanghai, China

Purpose:To compare the DWI image quality in BLADE and single-shot EPI sequence in sellar region. Methods: 3 healthy volunteers and 52 patients with sellar lesions participated in this prospective intra-individual study. Scanings were performed at 3.0T with BLADE DWI and single-shot EPI DWI sequence. SNR and image quality were evaluated. Lesions` sizes were measured for detecting their relationship with the image scores. Results: SNR showed no difference between two sequence and BLADE overperformed in image quality. The minimal lesion`s size for visualization was 12mm in BLADE and 28mm in EPI DWI. ConclusionBLADE DWI MRI can reduce artifacts and enhance the image quality in sellar region.