SNR Measurement on Single Images of Phantom Using Wavelet Transforms

Zhiyue J. Wang 1,2 and Jonathan M. Chia 3

1 University of Texas Southwestern Medical Center, Dallas, Texas, United States, 2 Children's Medical Center Dallas, Dallas, Texas, United States, 3 Philips Healthcare, Ohio, United States

In SNR assessment of phantom MR images, the noise level cannot be obtained from the background areas reliably when a parallel imaging technique is used. The current gold standard is to use a difference image for noise assessment on the object. In practice the difference image methods are prone to artifacts from position drifts and solution flow. We present a method for SNR measurement from one image only, based on the wavelet transform. The new method is validated using the difference image method.