More accurate estimation in prostate volume achieved by optimized single slab 3D SPACE compared with 2D-TSE: A quantitative study based on the radical prostatectomy specimens.

Souichirou Tateishi 1 , Masato Uchikochi 2 , Shohei Miyazaki 1 , Hitoshi Matsui 1 , Takashi Horinouchi 1 , Yasuhiko Tomita 3 , Kazuo Nishimura 4 , and Katsuyuki Nakanishi 1

1 Radiology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka, Osaka, Japan, 2 SIEMENS JAPAN K.K., Tokyo, Japan, 3 Pathology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka, Japan, 4 Urology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka, Japan

More accurate estimation of prostate volume is important for therapeutic trials. A recently evolved 3D-TSE T2WI sequence termed SPACE provides a high spatial resolution and isotropic volume image compared with the conventional 2D-TSE sequences. These improvements expect to provide more accuracy in prostate volume estimation. We compared 3D-SPACE with 2D-TSE about prostate volumes. Reference standard of prostate volume was obtained from pathology report as a gold standard. 3D-SPACE showed a higher correlation (SPACE; r=0.98, 2D-TSE; r=0.96) and a less relative error (p<0.01): (SPACE; 0.12}0.13, 2D-TSE; 0.19}0.15) with reference standard of prostate volume.