Is voxel-based apparent diffusion coefficient reproducible?

Masamitsu Hatakenaka1, Koichi Onodera1, Naomi Koyama1, and Mitsuhiro Nakanishi2

1Diagnostic Radiology, Sapporo Medical University, Sapporo, Japan, 2Division of Radiology, Sapporo Medical University Hospital, Sapporo, Japan

To evaluate reproducibility of voxel-based ADC, voxel-based ADC of the phantom was measured using clinical 3T MRI system. The direction of motion probing gradient affected the voxel-based ADC significantly in both echo planar imaging and turbo spin echo diffusion-weighted imaging. Also voxel-based ADC differed both between identical positioning examinations and between slightly different positioning examinations. Voxel-based ADC could not be reproduced sufficiently even in a phantom study. It would be recommended to pay enough attention when performing voxel-based ADC study like histogram analysis for tumor ADC.