Comparison of Image Quality and Apparent Diffusion Coefficient Reproducibility with Water Exitation Using Binominal Scheme 11, 121, 1331, 14641 in iShim Sequence

Xiaolu Li1, Mengchao Zhang1, Hong Zeng1, and Lin Liu1

1Jilin University Sino-Japan Hospital, Chang Chun, China, People's Republic of

Our study is to compare the homogeneity and degree of fat saturaton when using water excitation at different binominal scheme (11, 121, 1331, 14641) in iShim (sequence integrated shimming) DWI of liver.We found that the binominal scheme 1331 and 14641 in water excitation can improve the homogeneity of fat saturation in abdominal iShim DWI without effecting ADC values.We believe that it will be good for the clinical image quality.