10 fold accelerated 4D flow in the carotid arteries at high spatiotemporal resolution in 7 minutes using a novel 15 channel coil

Eva S. Peper1, Qinwei Zhang1, Bram F. Coolen2, Wouter V. Potters1, Pim van Ooij1, Dennis W.J. Klomp3, and Aart J. Nederveen1

1Radiology, AMC, Amsterdam, Netherlands, 2Biomedical Engineering and Physics, AMC, Amsterdam, Netherlands, 3Radiology, UMCU, Utrecht, Netherlands

Using a novel 15 channel coil for carotid artery imaging we could prove that the g-factor loss at higher acceleration factors is limited. This permits the use of higher parallel imaging factors than commonly exploited in carotid MRI. A 4D flow scan at high spatiotemporal resolution was accelerated 10 fold (SENSE 4 x 2.5), resulting in high quality images and consistent flow values acquired in a scantime as short as 7 min.