2D acquisition mode for T1 and T2 estimation using an ellipse-fitting approach on phase cycled bSSFP data

Yulia Shcherbakova1, Cornelis A.T. van den Berg2, Pim T.S. Borman1, Chrit T.W. Moonen1, and Lambertus W. Bartels3

1Center for Imaging Sciences/Imaging Division, University Medical Center Utrecht, Utrecht, Netherlands, 2Dept. of Radiotherapy/Imaging Division, University Medical Center Utrecht, Utrecht, Netherlands, 3Image Sciences Institute/dept. of Radiology, University Medical Center Utrecht, Utrecht, Netherlands

The ellipse fitting approach for simultaneous estimation of the relaxation times T1 and T2 from phase-cycled balanced steady-state free precession (PC-bSSFP) has been so far limited to the 3D acquisition mode due to sensitivity to slice profile imperfections. In this work we present the results of a 2D approach which is based on a RF excitation pulse optimization. This minimally affects the ellipse fitting leading to minimal corruption of T1 and T2 quantification from 2D PC-bSSFP data.