Better Depiction of Cystic Duct with Breath-hold 3D MRCP accelerated with GRASE as compared with FSE-based MRCP with Compressed Sensing at 1.5T

Mamoru Takahashi1, Yasuo Takehara2, Norihiro Tooyama1, Katsutoshi Ichijo1, Yasutomo Katsumata3, Akira Suwa3, and Harumi Sakahara4

1Seirei Mikatahara General Hospital, Hamamatsu, Japan, 2Nagoya University, Graduate School of Medicine, Nagoya, Japan, 3Philips Japan, Ltd, Tokyo, Japan, 4Hamamatsu Univ. Hospital, Hamamatsu, Japan

We compared the depiction of the cystic duct between two types of 3D breath-holding MRCP accelerated with GRASE (GRASE-MRCP) and FSE with compressed sensing (CS-MRCP) at 1.5T. Although imaging time and overall image quality was the same, breath-hold MRCP accelerated with GRASE allowed better depiction of the cystic duct as compared with FSE MRCP with CS because shorter TE with GRASE allowed higher signal of the concentrated bile.