Diffusion MRI of the Brain with a Gradient Strength of 100 mT/m and a Slew Rate of 1200 T/m/s

Bertram Wilm1, Manuela Roesler1, Franciszek Hennel 1, Markus Weiger1, and Klaas Pruessmann1

1Institute for Biomedical Engineering, University of Zurich and ETH Zurich, Zurich, Switzerland

The performance of diffusion-weighted single-shot EPI is limited by off-resonance artifacts and low signal-to-noise. To address both problems, we employ a recently developed gradient insert (strength: 100 mT/m, slew rate: 1200 T/m/s). Thereby the time for the diffusion encoding as well as the EPI readout train duration can be significantly shortened, resulting in higher signal yield and robustness against off-resonance artifacts. First in-vivo results are presented.