Meeting Banner
Abstract #3186

Low latency vs. non linear phase: A balanced approach for UNFOLD filter design in quantitative Real-Time MRI

Moonen C, Maclair G, Ries M
IMF CNRS / University of Bordeaux 2

Real-time control of interventional procedures such as guidance of catheters or ablation therapy require a high temporal resolution, a low processing latency and methods that preserve the temporal characteristics of the data. Although UNFOLD encoded sequences are interesting candidates for these applications, current filter implementations show either a non-linear phase (IIRs) which alters the temporal characteristics or have insufficient latency (FIRs). This study proposes an IIR filter design which satisfies both constraints of low latency and linear phase.

Keywords