Joshua FP van Amerom1, Michael S. Jansz1,
Christopher K. Macgowan1
1Departments of Medical Biophysics and
Medical Imaging,
Steady
state free precession phase-contrast (PC-SSFP) is proposed for measurement of
nontriggered time-averaged velocity as it exhibits less signal variation over
the cardiac cycle than conventional PC-MR. It is shown that Cartesian
nontriggered PC-SSFP provides an accurate mean-velocity measurement while
conventional nontriggered PC-MR overestimates the mean velocity due to
in-flow effects.
Keywords