We
  have verified thanks to a specific common quality control that 7 MRI devices
  included in a multicenter clinical project, had homogenous and acceptable
  characteristics to allow quantification and comparison between parameters
  extracted from different patient images acquired on different sites and to
  permit correlation with biopsies. This quality control was established with
  sequences from the MRI protocol. The studied parameters slightly varied
  depending on the different sites and MRI manufacturers and were in the
  awaiting of the project. This kind of quality control procedure should be
  included at the early beginning of any multicenter clinical projects
  involving quantitative MRI.