Matt Smith1,
1Medical Physics, University of
Wisconsin, Madison, WI, United States; 2Radiology, University of
Wisconsin, Madison, WI, United States; 3Biomedical Engineering,
University of Wisconsin, Madison, WI, United States
Regardless
of the sampling technique, the volume of interest is degraded by excited
signal energy outside the FOV. With 3D non-Cartesian acquisitions, the effect
is incoherent streaking with noise-like appearance. Commercial receive only
breast coils require slice/slab excitation from the body coil, exciting
tissue outside the FOV. A local transmit/receive breast coil based on a
solenoid design is compared with a commercial receive-only coil to
demonstrate that local excitation minimizes the unwanted signal energy
contaminating the FOV for non-Cartesian breast MR. Measurements of contrast
are higher and more consistent using a 3DPR SSFP sequence with a local
transmit/receive breast coil.
Keywords