Jeroen Cornelis Siero1, Marielle E.
  Philippens2, Arjan Willem Simonetti3, Johannes Marinus
  Hoogduin1, Peter R. Luijten4
1Brain Division, University Medical
  Center Utrecht, Utrecht, Netherlands; 2Radiotherapy, University
  Medical Center Utrecht; 3Philips Healthcare, Best, Netherlands; 4Radiology,
  University Medical Center Utrecht, Utrecht, Netherlands
The
  potential of cost function guided shimming was shown at 3T for finding
  optimal shim fields that minimize B0 inhomogeneities on a
  user-defined region of interest while confining the B0
  inhomogeneities outside this ROI. Experiments and simulations using the hybrid
  shimming approach show the possibility to control frequency selective fat
  suppression in abdomen and pelvic imaging while maintaining good B0
  homogeneity in the region of interest.
Keywords