Using
different undersampling patterns for the non-calibration and calibration
echoes has been shown to improve SNR per unit time of Parallel Imaging
accelerated IDEAL reconstructions by up to 40%. The different acceleration factors and
k-space undersampling patterns result in different noise enhancement in the
non-calibration and calibration echoes. In this work the T2*-IDEAL
reconstruction is modified to include noise weighting and demonstrate that
SNR improves with the modified reconstruction. For 14.2 fold accelerated phantom data, an
11.9% increase in mean SNR for all phantoms and a maximum 27% increase in SNR
over a single phantom was measured.