Wolfgang
Renz1,2, Tomasz Lindel, 23, Matthias Dieringer2,4,
Frank Seifert, 23, Jeanette Schulz-Menger, 2,5, Thoralf
Niendorf2,4, Bernd Ittermann, 23
1Siemens AG, Erlangen, Germany; 2Berlin
Ultrahigh Field Facility, Max-Delbrueck-Center for Molecular Medicine,
Berlin, Germany; 3Physikalisch-Technische Bundesanstalt (PTB),
Berlin, Germany; 4Experimental and Clinical Research Center,
Charite Campus Buch, Berlin, Germany; 5CMR-Unit, Charite Campus
Buch, Berlin, Germany
The
first steps in cardiac/body imaging at 7T have been reported. One of the
challenges is a suitable TX coil concept, which adresses the RF problems (B1
homogeneity, SAR) of body imaging at 300MHz. Traditional bodycoils seem not to
be the right way, a coil array enabling TX SENSE and B1 shimming seems to be
more promising. We describe a 8-element TX/RX loop coil array with adjustable
capacitive decoupling. A prototype has been realised and tested. First
imaging results in cardiac imaging are shown.
Keywords