Mikhail Kozlov1, Robert Turner
1Neurophysics, Max Planck Institute for
Human Cognitive and Brain Sciences,
We
investigated effects of head size and position on SAR for a commercially
available Rapid BioMed 7 T 8-element head coil. For this coil axial rotation
of the head can be considered safe, if the distance to lumped capacitors is
more than 20 mm. It is more dangerous to use this coil with the head only
partly inserted. The total head SAR
should be considered as the important safety limit, because the 3.2 W/kg
whole head SAR limit is reached sooner than the 10W/kg local SAR limit.