Chris A. Cocosco1, Andrew J. Dewdney2,
P Dietz2, M Semmler2, Anna M. Welz1, Daniel
Gallichan1, Hans Weber1, Gerrit Schultz1,
Juergen Hennig1, Maxim Zaitsev1
1Dept of Diagnostic Radiology, Medical
Physics, University Hospital Freiburg, Freiburg, B.W., Germany; 2Siemens
Healthcare, Erlangen, Germany
We
present the design considerations and evaluation measurements for the safety
of a PatLoc (Parallel Acquisition Technique using Localized Gradients)
gradient insert coil designed for human head imaging on a 3T MRI system. This novel concept has the potential to
allow higher gradient switching rates while not exceeding the Peripheral
Nerve Stimulation (PNS) limits. Based on the presented experimental
measurements and simulations, we consider imaging human volunteers with this
system to be safe.