Nicolas
Boulant1, Martijn Cloos1, Alexis Amadon1
1NeuroSpin, CEA, Gif sur Yvette, France
Efficient
mitigation of the radiofrequency inhomogeneity at high field using coil
arrays relies on the accurate knowledge of the individual B1 maps. To date,
no simple recipe has been formulated to correct for B0 inhomogeneity in the
evaluation of the B1 maps themselves. Here we derive a simple analytical
approximation to increase the accuracy of the B1 mapping techniques which
rely on the measurement of the flip angle using non-selective square pulses,
in the presence of B0 variations. In some possibly encountered cases,
applying the correction reduces the error of the estimated B1 amplitude from
13 % to 0.2 %.