Meeting Banner
Abstract #5057

Robust Method for EPI Ghost Correction

Frank Godenschweger1, Myung-Ho In1, Oliver Speck1

1Biomedical Magnetic Resonance, Institute for Experimental Physics, Otto-von-Guericke University, Magdeburg, Germany


A Nyquist ghost correction of EPI is propose, which determines the phase correction differences between channels and slices with high accuracy in a preparation step from a set of navigator echoes. Taking this calibration into account, only one single correction needs to be determined dynamically for all slices and channels during the EPI series, greatly improving stability.

Keywords