Inhomogeneities of the main magnetic field are known to cause considerable geometric distortion of EPI-acquired data. Accelerated acquisition reduces the problem, but does not eliminate it. We calculate field maps directly from the phase maps of the first two echoes in a multi echo EPI acquisition and use these to correct the residual distortions. The resulting geometrical fidelity yields a significant improvement in the co-registration to undistorted data.