Meeting Banner
Abstract #2727

Remotely Measuring Induced Radiofrequency Currents on Wires in MRI

Greg Hungtingdon Griffin1, Kevan J. Anderson2, Graham A. Wright, 12

1Medical Biophysics, University of Toronto, Toronto, Ontario, Canada; 2Sunnybrook Research Institute, Toronto, Ontario, Canada


Currently, safety of interventional devices is investigated using long experiments and direct measuring of induced temperature rises. A method of remotely measuring induced radiofrequency current using phase information from MRI is presented. Agreement between measurements and simulations is shown demonstrating accuracy of the remote method. This method presents a novel technique to rapidly investigate safety of different devices In Vitro.

Keywords