Abstract #3332

Analysis of BOLD Sensitivity in Matched Filter FMRI

Lars Kasper1, Maximilian Haeberlin1, Benjamin E. Dietrich2, Simon Gross1, Christoph Barmet2, 3, Christian C. Ruff4, Klaas E. Stephan1, 5, Klaas P. Pruessmann1

1Institute for Biomedical Engineering, University & ETH Zurich, Zurich, Switzerland; 2Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland; 3Skope Magnetic Resonance Technologies, Zurich, Switzerland; 4Department of Economics, University of Zurich, Zurich, Switzerland; 5Wellcome Trust Centre for Neuroimaging, University College of London, London, United Kingdom


Matched-filter fMRI utilizes 2D-density weighted EPI to increase BOLD sensitivity. It matches acquisition density in k-space to the desired spatial response typically a Gaussian kernel. Theory predicts significant SNR benefits from such acquisitions, which are maximal when thermal noise dominates and expected to decrease with higher physiological noise contributions. We explore the validity of this argument for different regimes of physiological noise in the brain and report consistent and replicable SNR increases of 20-40 % compared to uniform EPI acquisitions. For task-based fMRI, we observe a consistent increase in BOLD-sensitivity of 30 % (average t-value), and show reproducibility both within and between subjects (N=4).