Meeting Banner
Abstract #3947

A 16 Channel Cardiac Array for Accelerated Hyperpolarised 13C Metabolic Imaging on Pigs at 3T

Titus Lanz1, Markus Durst2, 3, Matthias Mller1, Francesca Frijia4, Giulio Govanetti5, Luca Menichetti5, Massimo Lombardi4, Jan Henrik Ardenkjaer-Larsen6, 7, Rolf F. Schulte2

1Rapid Biomedical, Rimpar, Germany; 2GE Global Research, Munich, Germany; 3Technical University Munich, Munich, Germany; 4Fondazione G. Monasterio CNR-Regione Toscana, Pisa, Italy; 5Institute of Clinical Physiology of CNR, Pisa, Italy; 6GE Healthcare, Copenhagen, Denmark; 7Danish Technical University, Copenhagen, Denmark


We show a setup for doing 13C imaging on hyperpolarised media in pigs on a clinical 3 T MR system. A 16 channel 13C Rx array is combined with a Tx birdcage, allowing the MR system's body coil for 1H imaging. Due to the low frequency the array is not sample noise dominated, but allows accelerated 13C imaging which is essential for efficient usage of the hyperpolarised magnetisation. First results of metabolic imaging on the pig heart are shown.

Keywords