Meeting Banner
Abstract #0091

Fast SEMAC by separation of on-resonance and off-resonance signals

Daehyun Yoon 1 , Valentina Taviani 1 , Pauline Worters 2 , and Brian Hargreaves 1

1 Stanford University, Palo Alto, CA, United States, 2 GE Healthcare, Menlo Park, CA, United States

We present a novel acquisition and reconstruction method to accelerate the Slice Encoding for Metal Artifact Correction (SEMAC) sequence for MR imaging near metallic implants. SEMAC adopted an additional encoding for slice-select dimension to resolve the off-resonance induced slice distortion, which severely increased the total scan time. In this abstract, we present a fast undersampling scheme and a simple reconstruction algorithm exploiting that the support of the extreme off-resonance spins is spatially limited. Our approach is to acquire on-resonance spins using less slice directional encoding and then to apply reduced FOV acquisition and reconstruction for off-resonance spins.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords