3D MR Imaging with the Dynamic Multi-Coil Technique: DYNAMITE MRI

Christoph Juchem 1 , Omar M Nahhass 2 , Terence W Nixon 1 , and Robin A de Graaf 1

1 Diagnostic Radiology, Yale University, New Haven, CT, United States, 2 RWTH Aachen University, Aachen, Germany

To date, MRI is based on linear field gradients generated by dedicated X, Y and Z coils. Here, multi-slice Dynamic Multi-Coil Technique (DYNAMITE) MRI is presented in which all field gradients are provided by MC field modeling. The obtained image fidelity is identical to conventional MRI based on dedicated gradient coils. Comparable image quality is a milestone towards the establishment of purely MC-based MRI systems. With the gains of DYNAMITE shimming in performance and efficiency, the MC technology has the potential to replace spherical harmonic coil systems for specific applications.