Meeting Banner
Abstract #1285

Comparison of analog and digital transceiver systems for magnetic resonance imaging

Seitaro Hashimoto 1 , Katsumi Kose 1 , and Tomoyuki Haishi 2

1 University of Tsukuba, Tsukuba, Ibaraki, Japan, 2 MRTechnology Inc., Tsukuba, Ibaraki, Japan

Analog and digital transceivers for MRI systems were critically evaluated under identical experimental conditions. The MRI experiments were performed using a 4.74-Tesla vertical-bore superconducting magnet and a high-sensitivity gradient coil probe. 3D spin-echo images of a kumquat were acquired with and without using a gain-stepping scan technique to extend the dynamic range of the receiver systems. The acquired MR images clearly demonstrated that nearly identical image quality was obtained for both transceiver systems, but DC and ghosting artifacts were obtained for the analog transceiver. We therefore concluded that digital transceivers have several advantages over the analog transceivers.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords