Abstract #1323
An 8-Channel Integrated Balun Phased Array (IBPA) for Small Anatomical Features
Wolfgang Loew 1 , Randy O. Giaquinto 1 , Brynne Williams 1 , J. Matthew Lanier 1 , Christopher Ireland 1 , Ronald Pratt 1 , and Charles Dumoulin 1
1
Imaging Research Center, Cincinnati
Children's Hospital Medical Center, Cincinnati, Ohio,
United States
A 3 Tesla 8-channel receive phased array for small
anatomical features is presented in this abstract. The
array uses a novel coil design integrating a balun into
each coil element. Coupling and matching was assessed
through network analysis. High-resolution imaging
experiments were performed on a phantom and human
fingers confirming high image quality and proof of
concept.