Abstract #1326
Performance comparison of a form fitted coil array vs. a quadrature birdcage coil for 31 P MRS studies in the human calf at 7T
Sigrun Goluch 1,2 , Andre Kuehne 1,2 , Ewald Moser 1,2 , and Elmar Laistler 1,2
1
MR Center of Excellence, Medical University
of Vienna, Vienna, Vienna, Austria,
2
Center
for Medical Physics and Biomedical Engineering, Medical
University of Vienna, Vienna, Austria
This work compares the performance of a 3 channel
transceive
31
P
surface coil with a comparable sized 16-leg birdcage for
the application in
31
P
localized spectroscopy experiments in the human calf at
7T with 3D electromagnetic simulation. B
1
+
as
well as SNR calculation for a comprehensive
investigation is presented for both designs, showing
higher B
1
+
and
SNR in the regions of interest for the surface coil,
while as expected the birdcage exhibits better
homogeneity. Since the ROIs are small voxels the
advantage of better homogeneity is of minor importance
compared to high sensitivity in terms of B
1
+
and
SNR.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.