A system for calibrated measurements of RF electromagnetic fields inside a clinical MR scanner

Gerd Weidemann 1 , Isabela Frese 1 , Frank Seifert 1 , Antonino Mario Cassara' 1 , Werner Hoffmann 1 , and Bernd Ittermann 1

1 Physikalisch-Technische Bundesanstalt (PTB), Berlin, Germany

For accurate and complete SAR assessment of an RF coil all E and B field components are required. A system for the calibration of fiber optic time-domain E 1 and B 1 field sensors using an MR compatible TEM cell and the MR scanner itself was developed. The complex field amplitudes of E 1 , B 1 + and B 1 were reliably measured in an ASTM body phantom equipped with provisions for internal and external field sensors. Utilizing the body coil of a clinical 3 T scanner, the measurements are used to evaluate the accuracy of EMF simulations.