Meeting Banner
Abstract #1386

Measuring the Spatial Magnetic Field Gradients within a Scanner Bore

Karlene M Fraser 1,2 , Elizabeth Morris 3 , Jonathan Ashmore 4 , Stephen Wastling 2 , Ruth O'Gorman 4,5 , and Gareth Barker 2

1 Neuroimaging, Maudsley Hospital, London, United Kingdom, 2 Centre for Neuroimaging Sciences, King's College London, London, United Kingdom, 3 Medical Engineering and Physics, King's College Hospital, London, United Kingdom, 4 Neuroradiology, King's College Hospital, London, United Kingdom, 5 Centre for MR Research, University Children's Hospital, Zurich, Switzerland

Manufacturers of medical implants and devices classify some as MR conditional, and specify maximum spatial magnetic field gradient (MFG) restrictions in G/cm or T/m. Measurements of the MF within the bores of a 1.5T and 3.0T HD.x, and two 3.0T MR750 MR scanners were obtained using a THM1176 Hall probe. Maximum spatial MFG calculated for the 1.5T and each 3.0T scanner was respectively over 400 G/cm and 700 G/cm, and approximately 1.5 and 2 times less that reported by the scanner manufacturer. Had the manufacturers measurements been restricted to the patient accessible area, our measurements may have been more comparable.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords