Meeting Banner
Abstract #1398

Improvement and Evaluation in PET Performance of 16-Channel Receive Anterior Array Coils for 3T Simultaneous PET/MR Scanner

Yun-Jeong Stickle 1 , Jianhua Yu 2 , Tae-Young Yang 1 , Sahil Bhatia 1 , and Dmitriy V Londarskiy 1

1 GE Healthcare, Aurora, OH, United States, 2 GE Healthcare, Waukesha, WI, United States

The purpose of this study was to develop and evaluate the novel coil designs with new materials adopted to reduce the coil impact on PET image quality. We developed a new plastic-free 3T 16-channel flexible anterior array (AA) RF coil with V0 flammability rating and biocompatibility (PET/MR prototype AA2) for torso and cardiac imaging and validated the PET performances for a GE conventional AA coil and two improved PET/MR prototype coils. The results show a 48% and a 79% average sensitivity loss improvement on PET/MR prototype AA1 coil and PET/MR prototype AA2 coil, respectively, from GE conventional AA coil.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords