Simultaneous Acquisition of Water and Metabolites Using Multi-Coil Sensitivities for Proton Chemical Shift Imaging Thermometry
Toru Shirai 1 , Satoshi Hirata 1 , Yo Taniguchi 1 , Yoshihisa Soutome 2 , Yoshitaka Bito 2 , and Hisaaki Ochi 1
1
Central Reseach Laboratory, Hitachi, Ltd.,
Kokubunji, Tokyo, Japan,
2
MRI
System Division, Hitachi Medical Corporation, Kashiwa,
Chiba, Japan
We have applied a method that acquires simultaneously
the water and metabolite signals using multi-coil
sensitivities for proton chemical shift imaging (CSI)
thermometry. The method consists of measurement shifting
only the water signal on the image, and a data process
separating the water and metabolite signals using the
coil sensitivity difference. The results of a cooled
phantom experiments revealed that the method may be
useful for acquiring temperature map without increasing
scan time and contaminating the frequency modulation
sidebands caused by water signal.