Abstract #3897
3D kat ARC cine for evaluating cardiac function on 3T
Shigeo Okuda 1 , Yoshitake Yamada 1 , Akihiro Tanimoto 1 , Jun Fujita 2 , Motoaki Sano 2 , Keiichi Fukuda 2 , Sachio Kuribayashi 1 , Atsushi Nozaki 3 , and Peng Lai 4
1
Diagnostic Radiology, Keio University School
of Medicine, Tokyo, Japan,
2
Cardiology,
Keio University School of Medicine, Tokyo, Japan,
3
MR
Applications & Workflow, GE Healthcare, Hino, Japan,
4
MR
Applications & Workflow, GE Healthcare, Menlo Park, CA,
United States
Whole-heart 3D cine was obtained in 18 sec using kat ARC
technique with acceleration factor of 7.7, following
SSFP 2D cine of the left ventricular (LV) short axis in
21 patients on 3T. LV functional parameters including
EDV, ESV, SV, EF, Mass were compared between 2D and 3D
cine. Bland-Altman analysis revealed the
interchangeability of parameters between 2D and 3D cine,
and inter- and intra-observer variability were excellent
in both techniques. 3D kat ARC cine enables accurate
clinical evaluation of cardiac function within a single
breathhold.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.