1 Radiology, Radboud university medical center, Nijmegen, Netherlands, 2 MRSRL, Electrical Engineering, Stanford University, Stanford, CA, United States, 3 Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 4 Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, Germany

A 1 H-MRSI sequence with spectral-spatial refocusing pulses was evaluated for prostate applications at 7T. The combination of an external Tx array coil and an Rx endorectal coil allowed the use of these non-adiabatic pulses, while retaining sensitivity within the prostate. Well-resolved spectra in the prostate of patients were obtained, with some variation due to B 0 and B 1 inhomogeneities.