Meeting Banner
Abstract #4113

1 H MR spectroscopic imaging of the prostate at 7T using spectral-spatial pulses

Miriam W. Lagemaat 1 , Marnix C. Maas 1 , Adam B. Kerr 2 , Andreas K. Bitz 3,4 , Stephan Orzada 4 , Mark J. van Uden 1 , Eline K. Vos 1 , and Tom W.J. Scheenen 1,4

1 Radiology, Radboud university medical center, Nijmegen, Netherlands, 2 MRSRL, Electrical Engineering, Stanford University, Stanford, CA, United States, 3 Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 4 Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, Germany

A 1 H-MRSI sequence with spectral-spatial refocusing pulses was evaluated for prostate applications at 7T. The combination of an external Tx array coil and an Rx endorectal coil allowed the use of these non-adiabatic pulses, while retaining sensitivity within the prostate. Well-resolved spectra in the prostate of patients were obtained, with some variation due to B 0 and B 1 inhomogeneities.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords