Design, Construction and Initial Evaluation of a Folded Insertable Head Gradient Coil

Trevor Paul Wade 1,2 , Andrew Alejski 1 , Janos Bartha 1 , Dina Tsarapkina 1 , R. Scott Hinks 3 , Graeme C. McKinnon 3 , Brian K. Rutt 4 , and Charles A. McKenzie 2

1 Robarts Research Institute, The University of Western Ontario, London, Ontario, Canada, 2 Medical Biophysics, The University of Western Ontario, London, Ontario, Canada, 3 GE Healthcare, Waukesha, WI, United States, 4 Radiology, Stanford University, Stanford, CA, United States

A symmetric, folded gradient, was designed for human brain imaging, with a target gradient strength of 80 mT/m and slew rate of 2900 mT/m/s. This design should limit forces and torques leading to a quieter coil with better eddy-current performance. The ultra-short design allows for human brain imaging without the need for shoulder cut-outs. The insertable coil was built using litz wire for transverse axes and hollow copper for Z to facilitate cooling. It was successfully interfaced with a clinical scanner and its electrical and imaging performance was characterised.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords