Abstract #4871
MRI Gradient Field-Induced Heating and its Frequency Dependency for Different Materials
Juan D. Snchez 1 , Mahdi Abbasi 1 , Amin Douiri 2 , Morwan Choli 3 , Wolfgang Goertz 2 , and Gregor Schaefers 3
1
Dept. of Numerical Simulation, MR:comp GmbH,
Gelsenkirchen, NRW, Germany,
2
Test
Laboratory, MR:comp GmbH, Gelsenkirchen, NRW, Germany,
3
Dept.
of Consulting, MR:comp GmbH, Gelsenkirchen, NRW, Germany
This study evaluated the gradient-induced heating of a
pair of metallic plates (dimensions similar to
commercial medical implantable pulse generators IPGs)
made of different materials in order to determine a
suitable test procedure for the test of gradient induced
heating. From the obtained results, it is observed that
a worst-case exposure cannot be concluded only by a
particular set of gradient field conditions for all
different implants.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.