Design and Testing of a 9-Channel Phased Array Coil for Magnetic Resonance Guided Focused Ultrasound Treatment of the Abdominal Region in Large Animal Models.

Emilee Minalga 1 , Robb Merrill 1 , Allison Payne 1 , Nassir F. Marrouche 2 , Dennis L. Parker 1 , and J. Rock Hadley 1

1 Radiology, University of Utah, Salt Lake City, UT, United States, 2 Comprehensive Arrhythmia Research and Management (CARMA) Center, University of Utah, Salt Lake City, UT, United States

Coils that increase the SNR for MRgFUS treatments are very important. More signal-to-noise ratio can give better temperature measurements which increases the treatment efficacy and safety. This work describes the design and testing of a 9-channel radio-frequency coil array designed for abdominal magnetic resonance guided focused ultrasound research in large animal models.