Abstract #1815
An interface to connect a 16-channel transmit array to an 8-channel parallel transmit system
Shajan G 1 , Jens Hoffmann 1 , Klaus Scheffler 1,2 , and Rolf Pohmann 1
1
Max Planck Institute for Biological
Cybernetics, Tuebingen, Baden Wuerttemberg, Germany,
2
Department
of Biomedical Magnetic Resonance, University Hospital,
Tuebingen, Baden Wuerttemberg, Germany
For head sized volume transmit arrays at ultra-high
field, 16 coil elements arranged in 2x8 is beneficial
than eight coils arranged in 2x4 configuration. The aim
of this project is to develop and qualify an interface
to connect a 16-channel transmit array to an 8-channel
parallel transmit system. The interface module consists
of eight capacitive loaded miniaturized Wilkinson power
splitters. Each pTx channel drives two coil elements
through the interface module. Transmit (B1+) field maps
were acquired in both CP mode and CP2+ mode and compared
with the numerical model. The transmit array performance
closely followed the predicted results.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.