Meeting Banner
Abstract #2409

Ghost-correcting SENSE reconstruction for multi-band EPI

Franciszek Hennel 1 , Aline Seuwen 1 , Constantin von Deuster 1 , and Klaas P. Pruessmann 1

1 Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland

Separation of slices simultaneously acquired in multi-band EPI is problematic when the N/2 ghost needs different correction parameters for different slices. We present a straightforward manner of combining the slice-dependent ghost correction with multiband SENSE. The method inverts the sensitivity encoding equations for separately reconstructed even- and odd-echo data. It achieves a significant complexity reduction compared to the previously proposed k-space-based approach.