A 3 channel 31 P and 2 channel 1 H coil array for 31 P NMR in the visual cortex at 7 T

Sigrun Goluch 1,2 , Andre Kuehne 1,2 , Albrecht Ingo Schmid 1,2 , Ewald Moser 1,2 , and Elmar Laistler 1,2

1 MR Center of Excellence, Medical University of Vienna, Vienna, Austria, 2 Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria

This abstract presents the development of a 3 channel 31 P and 2 channel 1 H RF coil array for phosphorous spectroscopy in the human visual cortex at 7 T. The naturally low sensitivity of 31P spectroscopy requires very efficient RF hardware for a successful implementation of an experiment. The array was modeled and optimized with the help of 3D electromagnetic simulation and was build and tested on the bench with a network analyzer. Preliminary results were obtained inside the scanner (Siemens, ) with phantom and fruit measurements, showing good results.