Abstract #3303
Total Variance Constrained Electrical Properties Tomography Using a 16-channel Transceiver Array Coil at 7T
Yicun Wang 1 , Xiaotong Zhang 1 , Jiaen Liu 1 , Pierre-Francois Van de Moortele 2 , and Bin He 1,3
1
Department of Biomedical Engineering,
University of Minnesota, Minneapolis, Minnesota, United
States,
2
Center
for Magnetic Resonance Research, University of
Minnesota, Minneapolis, Minnesota, United States,
3
Institute
for Engineering in Medicine, University of Minnesota,
Minneapolis, Minnesota, United States
Electrical Properties Tomography (EPT) is able to
provide quantitative maps of electrical properties (EP)
using MRI, which holds promise in early cancer
diagnosis. Currently, the Helmholtz Equation-based EPT
approaches ignore gradient information on the tissue
boundary, causing severe EP reconstruction artifacts;
other studies require either a priori boundary
conditions that are subject to error, or dielectrical
padding that may cause difficulties in practical
settings. In this study, we proposed a total variance
promoting gradient-based algorithm that does not bear
the aforementioned limitations. Simulation and phantom
experiment at 7T have demonstrated feasibility of this
new approach.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.