Meeting Banner
Abstract #3737

Advanced Intrinsic Correction of System Delays for Radial Trajectories

Martin Krmer 1 and Jrgen R Reichenbach 1

1 Medical Physics Group, Institute of Diagnostic and Interventional Radiology, Jena University Hospital - Friedrich Schiller University Jena, Jena, Germany

To perform correction of axis dependent system/gradient delays for radial imaging we propose an improved algorithm which works by iteratively optimizing the recon-structed image until the local minima of a calibration function is reached. The calibration function is calculated from the reconstructed image magnitude and thus uses the actually measured radial raw data for correction, requiring no additional calibration data to be acquired. We demonstrate the performance of the algorithm for both phantom and in vivo measurements.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords