MRI compatible-3D localization system for Real-Time catheter navigation
Olivia Garandeau 1 , Maxime Bories 1 , Fabrice Marquet 1 , Remi Dubois 2 , Pierre Jais 3 , and Bruno Quesson 1
1
IHU Liryc/CRCTB Inserm U1045, University of
Bordeaux, Pessac, Aquitaine, France,
2
IHU
Liryc/CRCTB Inserm U1045, ESPCI Paris Tech, Pessac,
Aquitaine, France,
3
CHU
bordeaux, Pessac, Aquitaine, France
Catheter exploration, key procedure for cardiac
arrhythmia diagnosis and treatment, could benefit from
MRI guidance and procedure monitoring. In this work, we
present a homemade MR compatible 3D and real-time
catheter navigation device based on electric field
detection. The system was validated in vitro providing
catheter positions with a maximum error of 5.5 mm at 25
frames per second. In vivo feasibility was also
demonstrated in a swine model. Such work opens
perspectives for catheter guidance and 3D slice tracking
during MR exam enabling real-time visualization of the
targeted area and respiratory motion correction.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.