Abstract #4213
MR Imaging of Knee Implants Using SEMAC at 3T
TAO AI 1 , Panli Zuo 2 , Yiqi Hu 1 , Mathias Nittka 3 , and Liming Xia 1
1
Radiology, Tongji Hospital, Tongji Medical
College, Huazhong University of Science and Technology,
Wuhan, Hubei, China,
2
Siemens
Healthcare, MR Collaborations NE Asia, Beijing, China,
3
Siemens
Healthcare, Germany, Erlangen, Germany
In order to investigate the efficacy of SEMAC and high
bandwidth (high BW) techniques in metal artifact
reduction and clinical applications at 3T MR imaging,
tissue specimens with CoCr/Ti and OxZr/Ti implants and
ten patients with total knee replacements were imaged,
and the artifact size, distortion and image quality were
quantitatively and qualitatively assessed. The results
demonstrated that SEMAC allows for significant metal
artifact reduction and distortion correction at 3T MR
imaging of knee implants, as compared with the
conventional and increased readout bandwidth sequences.
With optimizations, SEMAC would be routinely chosen for
the clinical application of knee prosthesis.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.