Meeting Banner
Abstract #0046

Reducing Brain MRI Artifacts Caused by Ferromagnetic Orthodontic Appliances Using Permanent Magnets

Zhiyue J Wang1,2, Yong Jong Park1,2, Youngseob Seo1,2, Michael C Morriss1,2, and Nancy K Rollins1,2

1UT Southwestern Medical Center, Dallas, TX, United States, 2Children's Medical Center, Dallas, TX, United States

Stainless steel orthodontic appliances are commonly found in adolescents undergoing clinical brain MRI examinations. They cause severe magnetic susceptibility artifacts and failure to obtain diagnostic information from many MR techniques. The B0 shimming capability present on clinical MR scanners cannot remove these artifacts. We have constructed devices for the correction of these artifacts at 1.5 T using small pieces of permanent magnets mounted on intra-oral mouth guards or an extra-oral mouth-band. The magnetic field from the permanent magnets cancels the B0 inhomogeneity induced by ferromagnetic orthodontic appliances, resulting in drastic improvement of MR image quality.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords