Meeting Banner
Abstract #0170

Comparison of 3T whole body parallel transmit arrays based on measured data from full scale models

Eddy B Boskamp1, Saikat Saha1, Ricardo Becerra1, and Michael Edwards1

1Engineering, GE Healthcare, Waukesha, WI, United States

In this study we are comparing 16 channel TEM, 8 channel TEM and 8 loop array pTx body coils based on experimental data obtained from full scale whole body prototypes as opposed to only simulation. Besides SAR, efficiency and uniformity, there are additional criteria to include when selecting a body coil for parallel transmit. Examples are star intensity artifact, E fields that heat up cables and baluns, VSWR, and perturbation sensitivity, which may make it impossible to build a certain design given obtainable tolerances.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords