Meeting Banner
Abstract #1857

Auto-Calibrating Wave-CS for Motion-Robust Accelerated MRI

Feiyu Chen1, Tao Zhang1,2, Joseph Y. Cheng1,2, John M. Pauly1, and Shreyas S. Vasanawala2

1Electrical Engineering, Stanford University, Stanford, CA, United States, 2Radiology, Stanford University, Stanford, CA, United States

In this work, we propose a motion-robust auto-calibrating Wave-CS technique. This technique uses the wave-encoded center k-space and the known point-spread-function (PSF) of wave-encoding to reconstruct a Cartesian central k-space for calibration. The coil sensitivity maps are subsequently estimated with ESPIRiT for CS-SENSE reconstruction of under-sampled k-spaces. Results show this approach can reduce the motion artifacts and the aliasing artifacts due to sensitivity variations between the calibration and accelerated wave-encoded acquisitions.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords