In MRI, gradient coil switching generates eddy currents in surrounding conductors including intra-coil eddy currents in the surrounding gradient coils (crosstalk). In order to investigate the intra-coil eddy currents, a set of gradient coils with differing track widths has been designed for use in a cylindrical MRI system. It was found that the surrounding coils with wide tracks produced significantly larger eddy currents than the cold shield. It is therefore necessary to take into account the crosstalk between coils when evaluating coil performance. We found that the optimal track width to use in gradient coil design is around 14 mm.