Meeting Banner
Abstract #3626

Magnetic field stability and homogeneity of high-temperature superconducting magnet with REBCO tapes measured using dedicated field camera system

Shin-ichi Urayama1, Taizo Tosaka2, Hiroshi Miyazaki2, Yasumi Ootani2, So Noguchi3, Hiroshi Ueda4, Atsushi Ishiyama5, Shunji Nomura2, Tsutomu Kurusu2, and Hidenao Fukuyama1

1Kyoto University, Kyoto, Japan, 2Toshiba corporation, Tokyo, Japan, 3Hokkaido University, Sapporo, Japan, 4Osaka University, Osaka, Japan, 5Waseda University, Tokyo, Japan

Because of the rapid rise of helium price in recent years, we started a project targeting to produce high-temperature superconducting ultra-high-field MRI systems with REBCO (Rare-earth - Barium - Copper Oxide) tapes and as the first step, we developed a middle-size 1T-REBCO magnet and evaluated its magnetic field stability and homogeneity with dedicated field camera system with 16ch NMR probes. Inhomogeneity was up to 800 ppm and stability with linear trend correction was ±1 ppm, and the causes were investigated.

This abstract and the presentation materials are available to members only; a login is required.